Specification

 

Type:

Diode Laser system with microprocessor control, touch screen and graphic user interface. Soft tissue diode laser for oral surgical procedures with fiberoptic delivery operating in contact or non-contact regime.

Size:

18.4 cm (7.25 in) width, 11.4 cm (4.5 in) depth, and 16.5 cm (6.5 in) height

Weight:

~ 1.8 kg (4 lbs)

Power requirements:

AC 120V/60Hz or 240V/50Hz with power input switch, maximum power consumption not to exceed 100 W.  90-132 and 198-264 volts AC, 50/60Hz.

Environment:

Operating temperature range of 15º- 30ºC (59º-86ºF), 10-90% relative humidity with ambient atmospheric pressures of 700 hPa to 1060 hPa

Foot pedal:

 Wired 2.5 meters cord length, IP67 rating,

Emission state:

Continuous wave or pulsed, pulse duration 25 µs to 3 ms, timer exposure duration or unlimited (as long as foot pedal pressed) duration. Power stabilization or tip temperature stabilization modes (user selectable). Timer duration is selectable from 50 ms to 99.9 s.

Emitted radiation:

980 nm ± 10 nm, up to 25 W output peak power, pilot beam 650 nm, up to 5 mW power

Protective glasses:

Optical density > 5, with 980 nm

Connectors:

 SMA high power fiber optic connector for the delivery system.

Optical fiber:  

400 µm core diameter, numerical aperture 0.39, resulting in Nominal Optical Hazard Distance (NOHD) of 931 cm, Maximum Permissible Exposure (MPE) of 363 mW /cm²)

Reprocessing:

Steam sterilization for handpiece. Non-immersible cold (liquid) disinfection or sterilization for fiber. Fiberoptic connectors are not allowed for liquid ingress.

Applied Part Rating:

Type BF applied part with no electrical connection with the console